ECM4, ECM5, ECM6

Model →
Features / specs ↓
ECM 4 ECM 5 ECM 6
LCD Display 32 Character Alphanumeric LCD Module with blue backlit
Number of Keys 7 6
Real Time Clock (RTC) Not Provided Yes
for Setting Delay Time and Sampling Time
Settable Sampling Start Delay Time Settable Up to 99 Hrs. 59 Min. Programmable through RTC
Settable Sampling Time Settable Up to 99 Hrs. 59 Min. Upto 999 hrs. 59 mts.
Electronic Time Totalizer Range (Cumulative Sampling Time) Up to 99 Hrs. 59 Min. Upto 999 hrs. 59 mts.
Automatic Sampling
Start / Stop Control
Fully programmable electronic timer to automatically shut off system after preset time interval.
Additional sampling control facility Nil Timer stops on no flow or very low flow
Cyclic Mode Programming Nil
Flow Measurement Basis Flow is inferred from differential pressure (Δp) across an exhaust orifice
Δp measurement device Pressure sensor
Flow inferring technique Computed programmatically
Displayed Flow Range 0. 6 to 1.8m3/min
Flow Setting Range Not Available 0.6 to 1.8 M3 / Min Not Available
Digital Flow Display Provided
Flow Control Mode Not Available Automatic Closed Loop Feedback Not Available
Flow Control Accuracy Not Applicable 2% of set value Not Applicable
Volume Totalizer Range 9999 M3 Upto 9999 M3 STP Corrected ?
Temperature Measurement (for STP correction) Not available Solid State Temperature Sensor
Internal Memory for data storage Capable of storing data pertaining to 1 sample Capable of storing data pertaining to 10 samples
Parameters Stored in internal memory Sampling time; volume; minimum, maximum and average flow Sampling time; volume, minimum, maximum and average flow; the minimum & maximum ambient temperature
Addition information stored in internal memory Nil Average flow, Constant Flow Volume, Non Constant Flow Volume with respective time stamps Not available
Communication port(s) for PC interface
(PC functions as Host and ECM as device)
Not available
PC Interface software Not available
Safety Features Not available