Legacy Mid Range PM10 Samplers

Available Models : PEM–RDS 4, PEM–RDS 4NL, PEM–RDS 5, PEM–RDS 6, PEM–RDS 6NL

Polltech’s Legacy Mid Range PM10 Samplers justifiably boast of a number of features which remain unmatched as yet by any other indigenous manufacturer. In fact, these features put the models in the international league of High Volume Air Samplers. Take a look at these features and also at a sample of the report and circular chart of the recorded flow.

Advanced High Tech Samplers

Model Comparisons :

The various models differ from each other in the type of blower used and type of f the electronic control module employed as shown in table below

Model →
Component / Device ↓
PEM-RDS 4 PEM-RDS 4NL PEM-RDS 5 PEM-RDS 6 PEM-RDS 6NL
Compatible Blower BL 1 NL1 BT 2 BL 1 NL1
Electronic Control Module Type
ECM 4
ECM 5
ECM 6
Model →
Features / specs ↓
ECM 4
ECM 5
ECM 6
LCD Display
32 Character Alphanumeric LCD Module with blue backlit
Number of Keys 7 6
Real Time Clock (RTC) Not Provided Yes
for Setting Delay Time and Sampling Time
Settable Sampling Start Delay Time Settable Up to 99 Hrs. 59 Min. Programmable through RTC
Settable Sampling Time Settable Up to 99 Hrs. 59 Min. Upto 999 hrs. 59 mts.
Electronic Time Totalizer Range (Cumulative Sampling Time) Up to 99 Hrs. 59 Min. Upto 999 hrs. 59 mts.
Automatic Sampling
Start / Stop Control
Fully programmable electronic timer to automatically shut off system after preset time interval.
Additional sampling control facility Nil Timer stops on no flow or very low flow
Flow Measurement Basis Flow is inferred from differential pressure (Δp) across an exhaust orifice
Δp measurement device Pressure sensor
Flow inferring technique Computed programmatically
Displayed Flow Range 0. 2 to 1.8m3/min
Flow Setting Range Not Available 0.6 to 1.8 M3 / Min Not Available
Digital Flow Display Provided
Flow Control Mode Not Available Automatic Closed Loop Feedback Not Available
Flow Control Accuracy Not Applicable 2% of set value Not Applicable
Volume Totalizer Range 9999 M3 Upto 9999 M3 STP Corrected ?
Temperature Measurement (for STP correction) Not available Solid State Temperature Sensor
Internal Memory for data storage Capable of storing data pertaining to 1 sample Capable of storing data pertaining to 10 samples
Parameters Stored in internal memory Sampling time; volume; minimum, maximum and average flow Sampling time; volume, minimum, maximum and average flow; the minimum & maximum ambient temperature
Addition information stored in internal memory Nil Average flow, Constant Flow Volume, Non Constant Flow Volume with respective time stamps Not applicable
Data Transfer to Pen Drive through USB port Not available
PC Interface software Not available
Safety Features Not Available

Accuracy Specifications

Model PEM–RDS 4
(ECM 4)
PEM–RDS 4NL
(ECM 4)
PEM–RDS 5
(ECM 5)
PEM–RDS 6
(ECM 6)
PEM–RDS 6NL
(ECM 6)
Flow Measurement Accuracy ± 3% ± 2% -
Flow Control Accuracy - ± 2% -
Time Measurement +/- 1 minute over 24 hours

General Specifications - Common to all (PEM–RDS 4, PEM–RDS 4NL, PEM–RDS 5, PEM–RDS 6, PEM–RDS 6NL)

Filter Holder SS 304 to accommodate filter paper of 25 x 20 cm
Filter Media Whatman GF/A or EPM – 2000
Particulate Sampling For PM10 (i.e <10µ ± 1µ) using size selective inlet at 1.13M3/Min
For SPM (i.e. <100µ) with Gable Roof (optional)
Construction Material & Finish Sturdy Powder Coated Aluminum

Operating Environment

Power 230 V AC ± 10%, 50 Hz.( 110 V, 60 Hz Optional )
Temperature 5 to 50oC
Relative Humidity Up to 95% non condensing

Physical Characteristics

Model PEM–RDS 4 PEM–RDS 4NL PEM–RDS 5 PEM–RDS 6 PEM–RDS 6NL
Size
( all dimensions in mm)
51 x42 x 123 450 x 380 1100 450 x 380 x 850 510 x 420 x 1230 450 x 380x1100
Weight 42 Kg 45 Kg 22 Kg 42 Kg 45 Kg
>