Legacy Mid Range SPM Samplers
SPM HVAS is basically used for collecting SPM which comprises particles of AED less than 100µ. This is achieved by incorporating as the sampling inlet a Gable Roof, fabricated as per US EPA / CPCB specifications.
By replacing the Gable Roof with Size Selective Inlet, the samplers can also be used for collection of particles less than 10µ and gets transformed into a PM10 High Volume Sampler.

We manufacture a range of models of High Volume Air sampler to suit various applications and budgets. The models are distinguished from one another in respect of the technological sophistication in flow measurement and control and also in blower model employed for generation of flow.
- Flow is inferred from differential pressure (Δp) across an exhaust orifice
- Δp is measured by a pressure sensor
- Flow is computed programmatically
- Digital display of inferred flow is provided
- Digital display of volume is provided
PEM–HVS 4, PEM–HVS 4NL, PEM–HVS 5, PEM–HVS 6 & PEM–HVS 6NL are called legacy samplers because these models have been in existence for a long time and have been widely accepted in the industry. These samplers are called mid range because of they are equipped with more sophisticated electronics than our rudimentary low cost models but are less sophisticated in the functionality than our Advanced High Tech Models PEM–HVS 7, PEM–HVS 8, PEM–HVS 8NL, PEM–HVS 9.
Salient Features :
- Microprocessor based Electronic Control Module
- Alphanumeric 32 character LCD Module
- Electronic Timer Cum Time Totalizer
- Settable Delay & Sampling Time
- Continuous Duty Blowers
- Orifice Plate at exhaust complies with USEPA.
- Gaseous sampling attachment as add on for all models.
- Can be morphed to a PM10 sampler by replacement gable roof with SSI.
Available Models : PEM–HVS 4, PEM–HVS 4NL, PEM–HVS 5, PEM–HVS 6, PEM–HVS 6NL
Model Comparisons :
The various models differ from each other in the type of blower used and type of f the electronic control module employed as shown in table below
Model → Component / Device ↓ |
PEM-HVS 4 | PEM-HVS 4NL | PEM-HVS 5 | PEM-HVS 6 | PEM-HVS 6NL |
Compatible Blower | BL 1 | NL1 | BT 2 | BL 1 | NL1 |
Electronic Control Module Type | ECM 5 |
Model → Features / specs ↓ |
|||
LCD Display | |||
Number of Keys | 7 | 6 | |
Real Time Clock (RTC) | Not Provided | Yes for Setting Delay Time and Sampling Time |
|
Settable Sampling Start Delay Time | Settable Up to 99 Hrs. 59 Min. | Programmable through RTC | |
Settable Sampling Time | Settable Up to 99 Hrs. 59 Min. | Upto 999 hrs. 59 mts. | |
Electronic Time Totalizer Range (Cumulative Sampling Time) | Up to 99 Hrs. 59 Min. | Upto 999 hrs. 59 mts. | |
Automatic Sampling Start / Stop Control |
Fully programmable electronic timer to automatically shut off system after preset time interval. | ||
Additional sampling control facility | Nil | Timer stops on no flow or very low flow | |
Flow Measurement Basis | Flow is inferred from differential pressure (Δp) across an exhaust orifice | ||
Δp measurement device | Pressure sensor | ||
Flow inferring technique | Computed programmatically | ||
Displayed Flow Range | 0. 2 to 1.8m3/min | ||
Flow Setting Range | Not Available | 0.6 to 1.8 M3 / Min | Not Available |
Digital Flow Display | Provided | ||
Flow Control Mode | Not Available | Automatic Closed Loop Feedback | Not Available |
Flow Control Accuracy | Not Applicable | 2% of set value | Not Applicable |
Volume Totalizer Range | 9999 M3 | Upto 9999 M3 STP Corrected ? | |
Temperature Measurement (for STP correction) | Not available | Solid State Temperature Sensor | |
Internal Memory for data storage | Capable of storing data pertaining to 1 sample | Capable of storing data pertaining to 10 samples | |
Parameters Stored in internal memory | Sampling time; volume; minimum, maximum and average flow | Sampling time; volume, minimum, maximum and average flow; the minimum & maximum ambient temperature | |
Addition information stored in internal memory | Nil | Average flow, Constant Flow Volume, Non Constant Flow Volume with respective time stamps | Not applicable |
Data Transfer to Pen Drive through USB port | Not available | ||
PC Interface software | Not available | ||
Safety Features | Not Available |
Accuracy Specifications
Model | PEM–HVS 4 (ECM 4) |
PEM–HVS 4NL (ECM 4) |
PEM–HVS 5 (ECM 5) |
PEM–HVS 6 (ECM 6) |
PEM–HVS 6NL (ECM 6) |
Flow Measurement Accuracy | ± 3% | ± 2% | - | ||
Flow Control Accuracy | - | ± 2% | - | ||
Time Measurement | +/- 1 minute over 24 hours |
General Specifications - Common to all (PEM–HVS 4, PEM–HVS 4NL, PEM–HVS 5, PEM–HVS 6, PEM–HVS 6NL)
Filter Holder | SS 304 to accommodate filter paper of 25 x 20 cm |
Filter Media | Whatman GF/A or EPM – 2000 |
Particulate Sampling | For PM10 (i.e <10µ ± 1µ) using size selective inlet at 1.13M3/Min For SPM (i.e. <100µ) with Gable Roof (optional) |
Construction Material & Finish | Sturdy Powder Coated Aluminum |
Operating Environment
Power | 230 V AC ± 10%, 50 Hz.( 110 V, 60 Hz Optional ) |
Temperature | 5 to 50oC |
Relative Humidity | Up to 95% non condensing |
Physical Characteristics
Model | PEM–HVS 4 | PEM–HVS 4NL | PEM–HVS 5 | PEM–HVS 6 | PEM–HVS 6NL |
Size ( all dimensions in mm) |
51 x42 x 123 | 450 x 380 1100 | 450 x 380 x 850 | 510 x 420 x 1230 | 450 x 380x1100 |
Weight | 42 Kg | 45 Kg | 22 Kg | 42 Kg | 45 Kg |
Add-ons
Size Selective Inlet for PM10 Sampling
By replacing Gable Roof with Size Selective Inlet, this instrument can be transformed into a PM10 High Volume Sampler ie for lower size particulate concentrations , sampling of particulates < 10µ
Standard Ambient Gaseous Sampling Equipment
This basic add-on with icebox cooling allows the HVS to be utilized for sampling of upto four gaseous pollutants in addition to particulates which is the primary function of the equipment.
Thermoelectric Cooled Ambient Gaseous Sampling Equipment
This advanced add-on with thermo electric cooling allows the HVS to be utilized for sampling of upto four gaseous pollutants in addition to particulates which is the primary function of the equipment.
Consumables
Filters of various media and sizes, Chemical Reagent Kits for Gaseous Sampling Attachments
Related Products
You may also be interested in
- PM10 High Volume Sampler
- Ambient Fine Dust Sampler
- Low Flow PM10/2.5 Sampler
- PUF Samplers
- Personal Air Sampler
- Dust Fall Jar